+GF+
Georg Fischer Signet
Home Sitemap About Us Deutsch
What's New   Applications   Products   Engineering Info   Sales   Support  
Signet Tech-Tips
Tech-Tip Index
 
Archive
Tech-Tip Reference # : TT_14_4_20mA_loop_signals Date Posted :
6/13/2006

4-20mA Loop Signals

A 4-20 mA signal is by far the most common method of transferring instrument signals to a PLC or chart recorder. When scaled, the 4 mA is equivalent to the low end of the range and the 20 mA is equivalent to the full scale of the range. For example, let’s look at a 3-8850 conductivity transmitter that is monitoring a solution on a scale of 0 to 2000 μS. The 4mA set point would need to be set to 0 μS and the 20mA set point needs to set to 2000 μS. If the measured solution is 1000 μS (middle of the user’s scale), then the resulting current will be 12mA (middle of the 4-20mA scale).

Using this quick calculation can determine if the mA output of the instrument is correct:

   •  4-20 mA scale has 16 mA (20mA – 4mA = 16mA)
   •  16mA divided by 2000 μS (full scale) = .008mA per 1 μS
   •  Add the 4mA to the calculated value
   •  1μS = 4.008mA

Signet instruments and some blind sensors offer two options when selecting a device with a 4-20mA output:
A passive loop or an active 4-20 mA loop.

A passive loop:

Requires the end user to provide a power source (typically 24 volts DC). The installer MUST pay attention to the wiring during installation or a reverse polarity signal may occur (a negative mA signal).  Rule of thumb, the power supply should maintain its polarity as shown in figure 1, the positive terminal of the power supply is wired to the positive terminal of the Signet instrument, and the negative terminal of the power supply is wired to the negative terminal of the device that is to capture the mA signal.


An active loop:

The active 4-20 mA signals, offered on the 3-8900 as well as ProPoint™ Transmitters,do not require an external power supply. The instrument provides the loop power and the installer only has to pay attention to the polarity of the instruments that are wired together fig. 2.

Important Note : ProPoint™ instruments do not have an isolated 4-20 mA output. It is very important that an isolated power supply is used to supply power to each ProPoint™ instrument. If a ProPoint™ instrument is powered by a shared power source, the 4-20 mA output could be inaccurate due to interference from other instruments or equipment using the same power source.



  Related Product(s)
 

   •  5500 ProPoint™ Flow Monitor
   •  5600 ProPoint™ Batch Controller
   •  5700 ProPoint™ pH/ORP Monitor
   •  5800CR ProPoint™ Conductivity/Resistivity Monitor
   •  5900 ProPoint™ Salinity Monitor
   •  8250 ProcessPro® Level Transmitter
   •  8350 ProcessPro® Temperature Transmitter
   •  8450 ProcessPro® Pressure Transmitter
   •  8550 ProcessPro® Flow Transmitter
   •  8750 ProcessPro® pH/ORP Transmitter
   •  8850 ProcessPro® Conductivity/Resistivity Transmitter
   •  8860 ProcessPro® Conductivity/Resistivity Controller
   •  8900 ProcessPro® Multi-Parameter Controller
   •  2350 Temperature Sensor (Blind)
   •  2450 Pressure Sensor (Blind)
   •  2551 Magmeter Flow Sensor
   •  2551 Magmeter Flow Sensor with Display
   •  2750 pH/ORP Sensor
   •  2850 Conductivity/Resist Sensor
3401 Aerojet Avenue, El Monte, CA 91731-2882, U.S.A.
© 2007 Georg Fischer Signet LLC
  • For Sales, click HERE
  • For Technical Service and Support, click HERE